+ Column Pair, OxyGo 5 Setting Sieve Columns - 05063420001
Skip Navigation Website Accessibility
Your Bridge to Independence
8987 Park Blvd N Seminole, FL
(727) 319-6756
 


Column Pair, OxyGo 5 Setting Sieve Columns

OxyGo Replacement Columns, Pair

Make sure you keep your OxyGo portable concentrator working at peak performance by replacing the sieve beds (columns) on a regular basis.  The easy-to-remove and install Columns are a quick and simple way to service your OxyGo at home, without the need for a repair shop.  Replacement Columns are sold as a pair, and should be replaced at the same time.  The column contains the sieve bed which converts room air into a high concentration of portable oxygen.  

Columns are a key component of your OxyGo and must be maintained to ensure proper oxygen output.   The manufacturer suggest replacing the columns about every 12 months and in certain operating conditions, earlier replacement may be required.

When the columns need to be replaced, you will receive an alert right on your OxyGo POC reading "O2 Service Soon". That is your indication that the columns should be replaced in the next 30 days.

Clear step-by-step instructions are available in your owner's manual.   On average, a consumer can change both columns in less than 10 minutes right in the comfort of your home.  No special tools required.  Your G3 offers you freedom and independence and you don't want to be without it - and with these easy to change Replacement Columns, you won't have to.

So simple, so inexpensive, and so vital to keep your POC performing at it's best!

Also fits Inogen G3.


Replacing Columns Instructions:

  1. Turn off the OxyGo concentrator by pressing the power button
  2. Remove the concentrator from the carry case, if applicable
  3. Remove the battery from the concentrator by pressing and holding the battery latch button and sliding the battery off
  4. Place the concentrator on its side so that the underside is visible. The metal column assembly can be seen on one side of the device
  5. Unlock the column assembly by pushing the latch button away from the columns
  6. While holding the button open, slide the column assembly out of the device by pulling on the column pull handle
  7. Remove the columns completely from the device, both columns are removed as one piece
  8. Remove the four dust caps, two at the top of the columns and two near the bottom, making sure there is no dust or debris where the dust caps were located
  9. As soon as the dust caps are removed, insert the column assembly into the OxyGo.
  10. Push the column assembly into the device such that the columns are fully seated into the OxyGo and the spring loaded latch button has fully returned to the closed position.

Once columns are replaced with new columns, reset the system software by performing the following reset operation:

  1. Turn off the concentrator and plug in external power.
  2. Hold the plus (+) and minus (-) buttons down for 5 seconds.
  3. When “Sieve Reset” appears on the LCD screen, press the mode button to confirm. This resets the system software.

Press the power button to resume normal use of the concentrator.